Canon (Toshiba) Infinix VF-i / Infinix 8000: X-ray tube end of life under heavy load
What happened
As reported in the case, on these busy angiography systems the X-ray tube simply wears out: one tube failed after a reported 1–2 years of round-the-clock three-shift operation, and an original tube was replaced after a reported five years at about fifteen procedures a day. No error codes were involved.
What was checked
- Listen to the anode rotor: worn bearings usually give themselves away by rumbling or grinding during rotation and a shortened spin-down time after the last exposure — an early warning long before a hard failure.
- Watch how the tube ages under load — arcing, unstable or faulted exposures, overheating complaints — and keep a simple log of procedure counts; the workload figures above are from this reported case, not a rule for every unit.
Cause
The engineers pointed to the rotating-anode bearings as the weak spot of this tube type: under continuous heavy load the bearings wear out ahead of the rest of the tube and effectively set its service life.
What was done
The worn tube was replaced (the part reported in the case was a DSRX-T7345GFS), followed by recommissioning checks per the manufacturer's service documentation. The case outcome was reported as solved.
Safety
Before any tube work, power down and lock out the system, verify the HV generator and cables are discharged (stored charge can be lethal), support the tube's weight during removal, and complete radiation output and electrical safety tests before the room returns to clinical use.
Lesson: On high-duty angiography systems the X-ray tube is a predictable wear part, so listen for bearing noise and log workload early enough to plan the replacement instead of waiting for a sudden failure.
Covered in the course: Angiography and C-arm Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.