Canon/Toshiba Xario/Aplio: unit completely dead, no indicators
What happened
The ultrasound scanner would not power up at all — no lamps or indicators lit anywhere on the unit. No error codes were shown; the unit looked completely dead, and the suspicion was that the mains contact fuses had tripped on their own.
What was checked
- With no indication at all, the input fuses were checked first — for continuity and visible burning.
- The engineers tried acting on the fuses mechanically to see if the trip window had shifted or could be reset by hand — this had no effect, as reported in the case.
- Then the unit was left unplugged from the mains for a couple of minutes and reconnected — this brought the power back, as reported in the case.
Cause
The mains contact fuses had tripped spontaneously. As noted in this case, on these Toshiba-built models the fuse elements are rated very close to the switch-on inrush current, so after years of power cycles they overheat, burn and degrade until they trip at a normal start-up — with no real fault in the scanner itself.
What was done
Disconnect the unit from the mains for a few minutes to let the fuses cool, then power it back up — this restored operation in the reported case. If a fuse has failed permanently, replace it with a new one of the same type and rating taken from the manufacturer's service documentation; never fit a higher rating or bridge it out.
Safety
Before opening the power supply or touching fuse holders, unplug the scanner from the wall outlet and give the internal capacitors time to discharge; after any fuse replacement, run the electrical safety tests before returning the unit to clinical use.
Lesson: On a completely dead ultrasound with no indicators, check the mains fuses first — on these models they are sized close to the inrush current and can trip simply from ageing, so a short cool-down or a like-for-like replacement often restores power without any real fault.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.