Carestream 5700: yellow deposit melting onto the print drum
What happened
A yellow deposit built up above the feed roller and, once heated, melted and dripped onto the print drum. A second printer of the same model, reported in the case for comparison, showed no such deposit.
What was checked
- Compared the affected printer with a healthy unit of the same model to confirm the deposit was not normal behaviour for this printer
- Inspected the print assembly and the engineers found a long-overdue, unchanged carbon filter and contamination around the roller area
Cause
Contamination of the print assembly: a carbon filter that had not been changed stopped capturing fumes, so deposits settled above the roller, softened with heat and dripped onto the drum.
What was done
The engineers cleaned the assembly with a suitable solvent, removed the left side cover, disassembled and cleaned the roller area, and replaced the carbon filter. The problem was solved and the printer returned to normal operation.
Safety
Work with the printer powered down and unplugged, ventilate the room when handling solvent, and run the electrical safety test before returning the unit to clinical use.
Lesson: When deposits appear inside an imaging printer, check consumables and filters first — a spent carbon filter can be the root cause, and comparison with a healthy unit quickly confirms it.