Carestream 5700: yellow deposit melting onto the print drum

What happened

A yellow deposit built up above the feed roller and, once heated, melted and dripped onto the print drum. A second printer of the same model, reported in the case for comparison, showed no such deposit.

What was checked

Cause

Contamination of the print assembly: a carbon filter that had not been changed stopped capturing fumes, so deposits settled above the roller, softened with heat and dripped onto the drum.

What was done

The engineers cleaned the assembly with a suitable solvent, removed the left side cover, disassembled and cleaned the roller area, and replaced the carbon filter. The problem was solved and the printer returned to normal operation.

Safety

Work with the printer powered down and unplugged, ventilate the room when handling solvent, and run the electrical safety test before returning the unit to clinical use.

Lesson: When deposits appear inside an imaging printer, check consumables and filters first — a spent carbon filter can be the root cause, and comparison with a healthy unit quickly confirms it.

Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.