Carestream · DryCr 30-x (первое поколение)

Carestream DryCr 30-x: banding on large-format prints only

What happened

The engineers reported streaky, overexposed bands on prints from large cassettes, while small formats printed clean. No error codes were logged and the built-in self-tests passed.

What was checked

Cause

The engineers found a brown deposit on the transport rollers — invisible to the eye and only showing up when wiped with a cloth. The deposit disturbed film travel and contact on wide formats but had little effect on narrow film.

What was done

The rollers were disassembled and cleaned; the banding disappeared completely on subsequent prints. In this case the fault was considered resolved.

Safety

Disconnect the imager from mains power and let the heated parts of the print engine cool before opening the film path; after reassembly, confirm all covers and interlocks are refitted and run a print-quality test before returning the unit to clinical use.

Lesson: When a film artifact appears on one format only while self-tests pass, wipe-test the transport rollers even if they look clean — deposits can stay invisible until you rub them.

Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.