Carestream DryCr 30-x: banding on large-format prints only
What happened
The engineers reported streaky, overexposed bands on prints from large cassettes, while small formats printed clean. No error codes were logged and the built-in self-tests passed.
What was checked
- The optical module and transport mechanisms were cleaned with a lint-free brush and wipes.
- Fabric tension on the optical module was compared with another unit of the same model.
- Cassettes were inspected: one new and one older cassette both showed the banding, so the media was ruled out.
- Inspection of the print matrix was recommended as a next step.
Cause
The engineers found a brown deposit on the transport rollers — invisible to the eye and only showing up when wiped with a cloth. The deposit disturbed film travel and contact on wide formats but had little effect on narrow film.
What was done
The rollers were disassembled and cleaned; the banding disappeared completely on subsequent prints. In this case the fault was considered resolved.
Safety
Disconnect the imager from mains power and let the heated parts of the print engine cool before opening the film path; after reassembly, confirm all covers and interlocks are refitted and run a print-quality test before returning the unit to clinical use.
Lesson: When a film artifact appears on one format only while self-tests pass, wipe-test the transport rollers even if they look clean — deposits can stay invisible until you rub them.
Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →