Carestream DryView 5700: film will not feed, button unresponsive
What happened
One case: a DryView 5700 imager, reported as about a year old with low print volume, refused to print. The control button either would not press or pressed with no reaction, film would not travel through the trays from any supply, and error 624 was reported in the case.
What was checked
- The engineers tried different film trays — the behaviour did not change, as reported in the case.
- They attempted to move the film manually along the feed path.
- Cleaning the densitometer was suggested in the discussion as a general step, but it did not address the feed problem.
- On opening the feed path, they found film wound around the feed roller under position 6 (position as reported in the case).
Cause
The engineers found film wound tightly around the feed roller under position 6 — a jam the printer could not clear by itself, so the feed shut down and the button appeared dead. Light usage plus film condition may have contributed, but that remains an assumption from one case.
What was done
They opened the film path, removed the wound film from the roller and cleared the path; printing returned to normal, with no parts replaced. Cleaning the densitometer and using fresh film were mentioned as general maintenance, but the actual fix in this case was unwinding the film from the roller.
Safety
Isolate the imager from the mains and let it finish its standby cycles before reaching into the film path — feed rollers and gears can pinch fingers — and run the electrical safety test before returning the unit to use.
Lesson: A dead button or a feed fault on a lightly used printer can still be a simple mechanical jam — inspect the film path for wrapped film before suspecting electronics or consumables.
Covered in the course: Medical Imagers and Film Printers — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.