Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Carestream Vita CR
Carestream · Vita CR

Carestream Vita CR: imaging plate corners creased when returning to the cassette

What happened

The reader folded the corners of the imaging plate while feeding it back into the cassette; the damage appeared to happen during the return stroke.

What was checked

Cause

The engineers found the feed rollers had degraded: they had lost their geometry and clamping force and turned sticky, so a roller pulled the plate up or down into the mechanism during the return stroke.

The cause is probable and was not fully confirmed.

What was done

The recommended fix in this case was to replace the cassette, straighten and re-seat the entrance shutter, and replace (or refurbish) the feed rollers. The engineers rated the outcome as probable, not yet confirmed.

Safety

Disconnect the reader from the mains and make sure the transport mechanism has fully stopped before opening it, and handle used plates and cassettes with gloves as patient-contact items.

Lesson: When plate corners crease on the return stroke, suspect degraded sticky feed rollers and a distorted cassette shutter rather than the plate itself, and check the whole feed path.

Covered in the course: X-ray Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes