CLIMA · MC15

CLIMA MC15: no reaction in CK-MB FS photometric assay

What happened

In the CK-MB FS photometric method the reaction did not go at all: neither on the CLIMA MC15 running in photometer mode, nor during calibration attempts on two Sapphire 400 analyzers. Reagents from two different lots gave the same result.

What was checked

Cause

Probable cause reported in the case: degradation of the CK-MB analyte in the serum. As colleagues noted, CK-MB is very labile — within half an hour to an hour it can fall to zero, and after thawing or dilution the serum stays usable for only a few minutes.

The cause is probable and was not fully confirmed.

What was done

No instrument repair was involved. The engineers recommended calibrating with freshly prepared serum and consulting the reagent manufacturer's chemists about analyte stability. The cause remained probable — it was not finally confirmed in the discussion.

Safety

Treat calibrators and serum as potentially infectious material: gloves and the lab's biosafety rules apply whenever you thaw, prepare or dilute samples.

Lesson: When an assay shows no reaction across several instruments and reagent lots, check sample and calibrator stability first — a labile analyte like CK-MB can be the culprit, not the analyzer.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.