Clima MS-15: program settings change on their own after mains dips
What happened
After power surges and dips, the analyzer started changing stored program settings by itself — as reported in the case, the incubation time of endpoint programs changed from 0 to 600 seconds. Operation was intermittent: some days results were correct, other days clearly wrong.
What was checked
- Checked how the settings are retained: the engineers found they are kept in battery-backed NVRAM modules (Dallas-type chips on sockets).
- Checked the internal supply voltages and currents; no fault was reported there.
- Re-checked and adjusted the photometer lamp.
- Re-checked reagents and the creatinine calibration/adaptation; this part was still under verification when the case was closed.
Cause
The probable cause in this case was poor mains quality: the room lights visibly flickered, and repeated power dips were corrupting the battery-backed settings memory and destabilising the instrument. The correlation was strong but not fully proven at the time of the note.
The cause is probable and was not fully confirmed.
What was done
The engineers installed a UPS with a 230 V output (as reported in the case), and operation improved — the settings stopped drifting. No board or memory modules were replaced; the socketed NVRAM chips remain the first place to look if the settings are lost again.
Safety
Unplug the analyzer from the mains and allow the power supply to discharge before working near the power board or the memory module, and complete the electrical safety checks before returning the unit to use.
Lesson: When stored settings start corrupting themselves, check mains quality and battery-backed memory early — a basic UPS can resolve symptoms that look like a firmware or chemistry problem.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.