Cormay Accent 200: low cholesterol results and failing cholesterol QC
What happened
The analyzer under-reported cholesterol on every sample, and the cholesterol test failed internal quality control, while all other tests passed QC normally. No error codes were reported in the case.
What was checked
- Check the cholesterol test parameters: the reagent blank optical density and the acceptable blank range — reported in the case as healthy when the blank is no higher than 0.05.
- Check the main wavelength of the test — reported in the case as 505, 515 or 546 nm — to rule out changes made in the test settings.
- Check the reagent itself: the engineers noted that non-working kits from this manufacturer had been seen before.
Cause
Not established in the case. Suspicion fell on the cholesterol test configuration (reagent blank OD limits set too tight or altered) or on a defective reagent kit; the engineers had not yet found the root cause — diagnostics were still in progress.
The cause is probable and was not fully confirmed.
What was done
The recommended step was to verify and, if needed, correct the standard test configuration — the reagent blank OD limits (reported in the case: lower bound down to -0.01, threshold 0.05) — without disabling the blank parameter check (a value of 0.00 switches it off and only hides the fault), then re-check the wavelength and rule out a bad reagent kit. Since the case was still in diagnostics, the outcome was probable, not confirmed.
Safety
Treat all patient samples and reagent spills as potentially infectious — wear gloves and disinfect surfaces — and before any work inside the analyzer, isolate it from the mains and let moving parts stop.
Lesson: When a single chemistry fails QC while the rest pass, start with that test's own settings and its reagent — and correct the cause instead of weakening or disabling the control.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →