Cormay · Clima 15

Cormay Clima 15: urea and creatinine calibration factors knocked out

What happened

On a benchtop chemistry analyzer the calibration factors for urea and creatinine were found knocked out of calibration, and there was no calibrator or control material on site to recalibrate properly; reported in the case.

What was checked

Cause

Reported in the case: the factors had been altered, apparently accidentally, during use, and the absence of calibrator and controls on site made an immediate proper recalibration impossible.

The cause is probable and was not fully confirmed.

What was done

The engineers performed an interim calibration against a sample with a known approximate value to restore operation, treated that setup as temporary, and scheduled full recalibration with the proper calibrator and controls as soon as they arrive; reported in the case, the outcome is probable — results issued before verification with controls should be regarded as approximate.

Safety

Treat patient samples, calibrators and any spilled reagent as potentially infectious — wear gloves and eye protection whenever you work on the cuvettes, sample path or fluid lines.

Lesson: A stopgap calibration only buys time — an analyzer must not keep reporting patient results without proper recalibration on calibrator and controls, and calibration settings should be protected and recorded so they cannot drift unnoticed.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.