CPI CPI200 IQ: generator switches off by itself

What happened

The x-ray generator (with its control monitor) began shutting down spontaneously during use, while the tube stand kept working normally.

What was checked

Cause

In this case the engineers found a power connector that had not been pushed fully home; over time the contact degraded and began dropping out, taking the generator down with it.

What was done

The poor contact in the power connector was located and corrected — the connector was fully re-seated and secured, after which the random shutdowns stopped (case reported as solved).

Safety

Before opening a generator cabinet, switch off and lock out the mains supply, allow the capacitors to discharge and verify zero voltage — the internal DC bus stores lethal energy.

Lesson: Always rule out a loose or half-seated connector before chasing a complex fault in generator electronics.