CPI CPI200 IQ: generator switches off by itself
What happened
The x-ray generator (with its control monitor) began shutting down spontaneously during use, while the tube stand kept working normally.
What was checked
- Check the generator's error log through its web interface to see whether the unit records why it lost power.
- Since the stand worked fine, focus on the generator's own power path: mains input, wiring and connectors.
- With the supply isolated, physically verify that every power connector is fully seated and locked.
Cause
In this case the engineers found a power connector that had not been pushed fully home; over time the contact degraded and began dropping out, taking the generator down with it.
What was done
The poor contact in the power connector was located and corrected — the connector was fully re-seated and secured, after which the random shutdowns stopped (case reported as solved).
Safety
Before opening a generator cabinet, switch off and lock out the mains supply, allow the capacitors to discharge and verify zero voltage — the internal DC bus stores lethal energy.
Lesson: Always rule out a loose or half-seated connector before chasing a complex fault in generator electronics.