Cryobatl · hc8e

Cryobatl hc8e: coolant leak at the plastic supply tube to the compressor

What happened

The unit was losing coolant through a weep at the plastic supply line going to the compressor. Re-tightening the fitting did not stop the leak.

What was checked

Cause

The plastic tube had hardened over time and cracked right at the seal area; tightening a cracked tube only chases the leak along the joint.

What was done

The cracked section was cut off and the swagelok-type joint (nut, conical seal, washer) was remade on fresh tube. In a similar reported situation, the whole tube was replaced while keeping the old nut and cone, which also cured the leak. The leak stopped after the repair.

Safety

Before opening any fitting on the coolant loop, switch the unit off and isolate power, let the circuit warm up and relieve residual pressure — trapped coolant can spray, and cold surfaces on a cryo system can cause frostbite.

Lesson: If re-tightening does not stop a leak on a plastic coolant line, the tube itself has usually hardened and cracked at the seal — plan on cutting back or replacing the tube, not just re-making the joint.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.