Cryomech Coolpak 6000: oil overtemperature trips after relocation
What happened
After the unit was relocated, the helium compressor began shutting down on oil overtemperature. Reported oil-sensor readings were 98-100 °C, and the cooling-water temperature delta rose above its earlier reported 5.5-6 °C; the trips stopped only when the helium working pressure was reduced to 18 bar.
What was checked
- Checked the cooling-water supply temperature and the delta: reported within 10-14 °C, so the incoming water itself was not the problem.
- Tried lowering the chiller setpoint and partly closing the bypass — the oil overtemperature trips continued.
- Measured helium temperatures at the cold head: reported supply 19 °C and return 23.8 °C, pointing to reduced heat removal on the water side.
- Suggested verifying actual water flow with a water meter rather than trusting chiller readings.
- Suggested starting with flushing the compressor's heat exchanger.
Cause
Most probable, as found in the case: fouling of the compressor's water-to-helium heat exchanger after the move — possibly sediment carried over from the chiller's expansion tank — restricting water flow and heat transfer. The raised water delta and the fact that the trips only stopped with a significant change in helium pressure point to a water-side cooling deficit, not a faulty oil sensor.
The cause is probable and was not fully confirmed.
What was done
Temporary measures reported in the case: helium working pressure raised to 20 bar and a fan directed at the compressor, after which the overtemperature trips stopped. The recommended proper fix is to start with flushing the cryocompressor's heat exchanger per the manufacturer's service documentation, then verify the flow and cleanliness of the whole water circuit.
Safety
Isolate the compressor from the mains and let it cool before opening panels or disturbing water or helium lines — the helium circuit holds stored pressure energy and the oil circuit runs hot; complete the manufacturer's electrical safety checks before returning the unit to use.
Lesson: After any relocation, suspect degraded cooling first — a fouled heat exchanger or changed water flow — when oil overtemperature appears; verify real flow and circuit cleanliness before blaming the sensor or the setpoints.
Covered in the course: Fault-Finding Method for Medical Equipment →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.