Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Dameca MRI 508
Dameca · MRI 508

Dameca MRI 508: measured tidal volume 200–300 mL low or erratic with a normal curve

What happened

In the reported case, with a set tidal volume of 610 mL the machine displayed values roughly 200–300 mL low or randomly off, while the flow curve on the screen looked correct. No error codes were present.

What was checked

Cause

The engineers concluded that the cause was most likely a restriction at the test lung's tee, creating turbulent flow that the flow sensor could not measure correctly; the machine itself was found healthy.

What was done

Verification on a plain reservoir bag showed the readings followed the test load, not the machine, and the extension piece in front of the test lung did not remove the deviation. The machine was returned to service and, as reported in the case, the procedure was completed successfully.

Safety

Before returning an anesthesia machine to use, complete the full system test and the pre-use leak and ventilation check with the exact circuit configuration that will be connected to the patient.

Lesson: When measured volume disagrees with the set value but the waveform looks normal, test the machine against a different passive load before replacing the flow sensor — the geometry of a test lung can distort the measurement.

Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes