Dameca · MRI 508

Dameca MRI 508: water in the IBS breathing block, leak and pressure overreading

What happened

Water kept collecting in the IBS breathing block of an anesthesia machine. It showed up as a leak and too-high pressure sensor readings, and during the test water dripped from the valves with the red rubber rings. Reported in the case as a problem since the machine was first put into use.

What was checked

Cause

Not a fault. The water is normal condensate: the reaction of the absorbent with CO2 releases water and heat, so moisture steadily accumulates in the breathing circuit during normal operation.

What was done

Nothing to repair or replace. The engineers advised periodic disassembly of the breathing block to drain the collected water, then drying and wiping the parts. The full cleaning is routine user maintenance, done by the user according to the manufacturer's instructions. With regular draining and drying, the leak and the pressure overreading were resolved.

Safety

Before opening the breathing block, disconnect the machine from power and gas supply and vent the circuit; handle spent absorbent with gloves as it can be warm and dusty, and run the leak test again before the machine returns to use.

Lesson: Steady water build-up in an anesthesia breathing block is usually normal absorber chemistry, so scheduled draining and drying prevents phantom leaks and false pressure readings.

Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.