Dameca · MRI508

Dameca MRI508: water standing in the air rotameter after pipeline connection

What happened

When the anesthesia machine was connected to the medical gas console, water came out of the air supply line and entered the machine, and a column of water stood in the air rotameter. The engineers found the machine had not been powered on, which limited the possible damage.

What was checked

Cause

Most probable: condensate had collected in the medical air pipeline, and connecting the machine let that water run into the air supply line and up into the rotameter. The exact source of the water was not confirmed in the case.

The cause is probable and was not fully confirmed.

What was done

As reported in the case: purge the machine with dry air from a known-good compressor at 5–8 L/min in manual mode, with the breathing circuit, absorber canister and bottle removed, for several hours. The outcome is probable, not confirmed; if moisture remains in the rotameter or the internal tubing after drying, the machine should not go back into service without further checks.

Safety

Water in the gas path of an anesthesia machine is a direct patient hazard: keep the machine off mains until fully dried, and before return to use perform a complete leak test and the electrical safety test.

Lesson: Before connecting a machine to a gas pipeline, verify the supply is dry — a wet line will fill anything you plug into it.

Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.