Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Datex-Ohmeda (GE) Aespire/Avance S5
Datex-Ohmeda (GE) · Aespire/Avance S5

Datex-Ohmeda Aespire/Avance S5: CAIO gas analyzer under-reads oxygen

What happened

With 100% oxygen selected on the anesthesia machine, the CAIO gas-analysis module displayed no more than 60% O2, as reported in the case.

What was checked

Cause

Probable cause: calibration drift of the CAIO gas-analysis module. Since the circuit was proven to carry 100% oxygen, the low reading pointed at the analyzer itself, not at the gas delivery.

The cause is probable and was not fully confirmed.

What was done

Calibrate the module with the special calibration gas specified by the manufacturer, following the procedure in the documentation; where that routine is restricted, it is a procedure from the manufacturer's service documentation, done by authorised personnel. At the time of the report the engineers were still waiting for a regulator for the calibration-gas cylinder, so the outcome is probable.

Safety

Treat the breathing-circuit sample line as contaminated with patient secretions (biohazard) when disconnecting it, and before return to clinical use verify the oxygen reading against a gas mixture of known concentration.

Lesson: When a gas analyzer disagrees with the delivered mixture, first prove which side is wrong by sampling the same gas on a known-good analyzer before working on the module.

Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes