Datex Ohmeda Aespire View: PEEP will not drop to the set level
What happened
As reported in the case, PEEP stayed at about 10 cmH2O and would not fall to the set value; swapping the flow sensors with another machine changed nothing.
What was checked
- Flow sensors were swapped from another machine of the same type — no change, so the sensors were ruled out.
- The pneumatic block was removed and disassembled for inspection and cleaning.
- Flow valve calibration was carried out after cleaning according to the manufacturer's service documentation.
Cause
In this case the probable cause was water entering the machine through the air supply line: liquid in the pneumatic block kept the flow valve from venting properly, so the circuit held pressure and PEEP would not fall.
The cause is probable and was not fully confirmed.
What was done
The engineers found that the way through was to pull the pneumatic block, disassemble and wash it out, then calibrate the flow valve and replace it if calibration does not restore control. The outcome was reported as probable, not fully confirmed.
Safety
Disconnect the machine from mains power and from all gas supplies and let the circuit fully depressurise before opening the pneumatic block, and after any work on the breathing gas path complete the full pre-use check and an electrical safety test before the machine returns to clinical use.
Lesson: A PEEP that will not fall points more to dirty or flooded pneumatics than to a bad sensor — trace the gas path and look for water ingress before swapping parts.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.