DGM Z-150: vacuum pump end-of-life under heavy daily use

What happened

A hydrogen peroxide plasma sterilizer reportedly lost vacuum performance after years of heavy daily cycling, and the original vacuum pump was suspected to be worn out.

What was checked

Cause

Not confirmed in the case; the picture is consistent with normal wear of the original rotary-vane pump after long service on hydrogen peroxide vapour.

What was done

The engineers replaced the pump with a new original TRIVAC D16 rotary-vane pump to the original specification, renewed the oil and filters, and returned the unit to service. As reported in the case, it has since run into its fourth year at a reported 10–14 cycles per day, needing only scheduled oil changes, filters and hydrogen peroxide top-ups.

Safety

Treat used pump oil, filters and empty peroxide cassettes as chemical and biohazard waste, and confirm chamber leak-tightness and a passed electrical safety test before the sterilizer returns to clinical use.

Lesson: A hydrogen peroxide sterilizer can carry a heavy daily load for years when the vacuum system stays on the approved pump with disciplined oil and filter changes.

Covered in the course: Steriliser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.