DGM Z-150: vacuum pump end-of-life under heavy daily use
What happened
A hydrogen peroxide plasma sterilizer reportedly lost vacuum performance after years of heavy daily cycling, and the original vacuum pump was suspected to be worn out.
What was checked
- Compare pump-down time and ultimate vacuum per cycle with the figures in the manufacturer's service documentation.
- Check pump oil level and condition; dark, emulsified or contaminated oil points to pump wear or peroxide ingress.
- Run a chamber and tubing leak (rate-of-rise) check before condemning the pump.
- In this case the engineers also reviewed the fault history in a remote session with manufacturer service, as reported.
Cause
Not confirmed in the case; the picture is consistent with normal wear of the original rotary-vane pump after long service on hydrogen peroxide vapour.
What was done
The engineers replaced the pump with a new original TRIVAC D16 rotary-vane pump to the original specification, renewed the oil and filters, and returned the unit to service. As reported in the case, it has since run into its fourth year at a reported 10–14 cycles per day, needing only scheduled oil changes, filters and hydrogen peroxide top-ups.
Safety
Treat used pump oil, filters and empty peroxide cassettes as chemical and biohazard waste, and confirm chamber leak-tightness and a passed electrical safety test before the sterilizer returns to clinical use.
Lesson: A hydrogen peroxide sterilizer can carry a heavy daily load for years when the vacuum system stays on the approved pump with disciplined oil and filter changes.
Covered in the course: Steriliser Service — Level 2: Fault-Finding →