Diamant Mammo-4MT: R02 fault during exposure
What happened
On a Diamant Mammo-4MT digital mammograph, error R02 appeared during exposure. In this case the engineers tied it to the filament board, which is sensitive to the incoming mains voltage; the problem was described as recurring seasonally, but this is one case, not a rule.
What was checked
- Measure the mains voltage on the phase feeding the unit; in the case, about 235 V was given as the upper level worth checking.
- If the line runs high, reconfigure the transformer wiring to the 240 V tap, following the manufacturer's service documentation.
- Measure the Vsafety reference on the filament board at test point TP3; in the case, a value below about 4 V was linked to the R02 error.
Cause
The filament board is sensitive to input voltage: when its Vsafety reference sags below about 4 V, the system reports R02. In this case the mains ran high enough to pull the reference down, which the engineers described as a seasonal pattern for this board.
The cause is probable and was not fully confirmed.
What was done
The mains supply was brought back in range, including moving the transformer to the 240 V tap per the service documentation, and the Vsafety reference was adjusted with potentiometer RV3 on the filament board until TP3 read about 5.4 V, as reported in the case. The outcome was considered probable: keep the unit under observation and confirm the error does not return.
Safety
Isolate the unit from the mains and let the storage capacitors discharge before touching the transformer wiring; while probing the live board, make sure an exposure cannot be triggered, and run the electrical safety checks before returning the unit to clinical use.
Lesson: Before blaming an analog board for intermittent exposure errors, measure the mains and the board's reference voltages — a plain supply-voltage problem can look like a board failure.
Covered in the course: Mammography Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.