Diatron · Abacus junior 30

Diatron Abacus junior 30: every result reads several times too high

What happened

Reported in the case: after a dispenser swap, the analyzer returned all results several times higher than expected, with no error codes. During a check dilution, a drop of blood stayed on the needle and the dilution looked brownish.

What was checked

Cause

The engineers found the instrument had been left in predilution mode while whole-blood samples were being run, so every measurement carried an extra dilution factor and all counts came out multiplied.

What was done

Setting the analyzer to whole-blood mode, matching the sample actually being fed, returned all results to normal; the case was closed as solved with no parts replaced.

Safety

Treat the needle, sampler and any spilled dilution as infectious blood: wear gloves, disinfect the sampling area, and confirm correct readings on control material before the analyzer returns to use.

Lesson: Before opening up the fluidics, check that the selected measurement mode matches the sample being fed — a wrong mode can imitate a gross dosing fault.