Diatron Abacus junior 30: every result reads several times too high
What happened
Reported in the case: after a dispenser swap, the analyzer returned all results several times higher than expected, with no error codes. During a check dilution, a drop of blood stayed on the needle and the dilution looked brownish.
What was checked
- Dispenser replaced to rule out a dosing error
- Valve before the sampling syringe checked
- Needle rinse after aspiration checked for carryover
- Measurement mode compared against the sample type (whole blood vs predilution)
Cause
The engineers found the instrument had been left in predilution mode while whole-blood samples were being run, so every measurement carried an extra dilution factor and all counts came out multiplied.
What was done
Setting the analyzer to whole-blood mode, matching the sample actually being fed, returned all results to normal; the case was closed as solved with no parts replaced.
Safety
Treat the needle, sampler and any spilled dilution as infectious blood: wear gloves, disinfect the sampling area, and confirm correct readings on control material before the analyzer returns to use.
Lesson: Before opening up the fluidics, check that the selected measurement mode matches the sample being fed — a wrong mode can imitate a gross dosing fault.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.