Dilon Diamond (Paxscan detector): detector link drops after a single image

What happened

On a digital X-ray room with a flat-panel detector, the detector link dropped periodically: no images reached the acquisition workstation. After each reboot the detector passed exactly one image, then went silent again.

What was checked

Cause

The engineers found a software cause, not hardware: the acquisition software stack did not persist the correct driver type for the grabber card, so after every reboot the card came up with the wrong driver and the detector link collapsed after the first acquisition.

What was done

The acquisition software was updated to the newer version reported in the case (version 6). After the update the network card was correctly identified as an Intel Pro 1000 grabber and image transfer became stable. No parts were replaced — a standard software update from the manufacturer was enough.

Safety

Isolate the system from the mains before checking or reseating the grabber card or detector power connections, and before returning the room to service confirm stable image transfer with repeated test exposures made only when the controlled area is clear.

Lesson: If a networked detector works for exactly one acquisition after a reboot, suspect the acquisition software and grabber driver stack first and bring them to the manufacturer's supported version before touching hardware.

Covered in the course: X-ray Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.