DIRUI CS-400: water bath sensor fault
What happened
The analyzer kept showing a water bath sensor problem. Replacing the sensor ribbon cables and the sensors themselves did not help, and outside specialists called in on the case could not find the cause either.
What was checked
- Replaced the sensor ribbon cables — the fault stayed.
- Replaced the water bath sensors themselves — no change.
- Inspected the sensor board in the white housing with the indicator LED and found it loose inside the cabinet, with no secure fixing.
Cause
The engineers found that the water bath sensor board (the module in the white housing with an LED) was poorly secured and hung loose inside the instrument, so its circuit behaved as if the sensor itself were faulty.
What was done
The board was firmly bolted to the instrument chassis. As reported in the case, the fault did not return — the analyzer has run without this problem for about five years since.
Safety
Disconnect the analyzer from the mains and let the water bath cool before working near the bath or its wiring, and repeat the electrical safety checks before returning the analyzer to use.
Lesson: When new sensors and cables do not clear a sensor fault, check how the sensor electronics are mounted — a loose board can look exactly like a bad sensor.