Academy / Field fault notes / DIRUI CS-400
DIRUI · CS-400

DIRUI CS-400: water bath sensor fault

What happened

The analyzer kept showing a water bath sensor problem. Replacing the sensor ribbon cables and the sensors themselves did not help, and outside specialists called in on the case could not find the cause either.

What was checked

Cause

The engineers found that the water bath sensor board (the module in the white housing with an LED) was poorly secured and hung loose inside the instrument, so its circuit behaved as if the sensor itself were faulty.

What was done

The board was firmly bolted to the instrument chassis. As reported in the case, the fault did not return — the analyzer has run without this problem for about five years since.

Safety

Disconnect the analyzer from the mains and let the water bath cool before working near the bath or its wiring, and repeat the electrical safety checks before returning the analyzer to use.

Lesson: When new sensors and cables do not clear a sensor fault, check how the sensor electronics are mounted — a loose board can look exactly like a bad sensor.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.