Draeger Evita 4: mixer fault with constant air pumping
What happened
The ventilator kept trying to pump air even in standby and reported a gas mixer fault. When supplied from a compressor, it indicated that no air pressure was present.
What was checked
- Checked the supply setup, with the machine running from a compressor instead of wall outlets.
- Looked for water carried from the compressor into the ventilator.
- Checked the compressor condensate drain valve.
Cause
Not finally confirmed. The most probable cause reported in the case is a failed air proportional valve, possibly due to water or condensate reaching the mixer through the air supply; oxidation at the air inlet port was also suspected.
The cause is probable and was not fully confirmed.
What was done
The engineers recommended checking the air proportional valve and removing the air connection port to inspect it for oxidation. In this case the cause remained probable, not confirmed.
Safety
Before opening the pneumatic path, disconnect the ventilator from the patient and shut off and depressurise the gas supplies; after any mixer or valve work, complete the manufacturer's verification and an electrical safety test before the unit returns to use.
Lesson: When a ventilator reports mixer faults, trace the supply side first, since moisture from an air compressor can reach and damage the mixer pneumatics.
Covered in the course: Ventilator Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.