Draeger · Savina

Draeger Savina: set tidal volume not reached at higher settings

What happened

As reported in the case, this Savina ventilator would not establish delivered tidal volumes above 330 mL, while settings below 300 mL worked adequately. Flow sensor calibration was checked and the membranes were found intact, so turbine wear was suspected.

What was checked

Cause

The cause was not confirmed in the case. The most probable explanation was contamination of the exhalation valve membrane: deposits on the membrane affect how the valve seats and moves, and this tends to show up first in the volume measurement at higher delivered volumes.

The cause is probable and was not fully confirmed.

What was done

The fix suggested in the case was to wash the exhalation valve membrane in warm water, following the manufacturer's cleaning instructions for that part, and to reassemble only after it had dried completely. The outcome was reported as probable, not verified by measurement, so a post-repair volume check is essential.

Safety

Isolate the ventilator from the patient and from mains power before removing the exhalation valve, and before return to use verify delivered volume and circuit leak-tightness on a test lung and complete the electrical safety test.

Lesson: On a tidal volume shortfall that appears only at higher settings, clean and inspect the exhalation valve membrane before suspecting the flow sensor or the turbine.

Covered in the course: Ventilator Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.