Dräger 7100/9100: suspected sticking flow valve
What happened
A flow valve fault on a Dräger anaesthesia machine, reported in the case as a 7100 or 9100 while the model was still being clarified. No specific symptoms were described; the picture suggested the valve was sticking rather than failing electrically.
What was checked
- Check whether liquid could have entered the machine through the air line and trace where it came from.
- Disassemble the pneumatic block and inspect it for fluid traces and oxidation, especially around the flow valve.
- On the 9100, run the full pre-use test before starting work — reported in the case, it returns an error when the flow valve is faulty.
Cause
Probable cause: the flow valve sticking after liquid entered the pneumatics through the air line and left oxidation on the valve parts. This remained a suspected, not confirmed, diagnosis.
The cause is probable and was not fully confirmed.
What was done
The fix in this case: open the pneumatic section, remove the fluid traces and oxidation and restore free movement of the flow valve. Reassembly, adjustment and final verification were a procedure from the manufacturer's service documentation, done by authorised personnel. The outcome was reported as probable, not confirmed.
Safety
Before opening the pneumatic block, isolate the machine from mains power and gas supplies and let it fully depressurise, and before it returns to clinical use run the full pre-use test with a leak and ventilation check.
Lesson: When a flow valve misbehaves on an anaesthesia machine, check the whole gas path for liquid ingress first and treat the full pre-use test as the final gate before the machine goes back to the patient.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.