Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Dräger Apollo (авент)
Dräger · Apollo (авент)

Dräger Apollo: sensors oxidised across all pneumatic channels

What happened

Suspected leak in the pneumatic block of a Dräger Apollo anesthesia machine; on inspection, the sensors in all pneumatic channels were found oxidised.

What was checked

Cause

Moisture had been getting into the pneumatic block over time — the machine was wiped down with a wet cloth — and gradually oxidised the sensors in all channels. Reported in the case as a known issue on these machines when liquid gets in, so all channels failing together pointed to one shared cause rather than separate sensor faults.

What was done

The case record does not detail the repair steps; the machine went back into service once the oxidised sensors in the pneumatic block had been dealt with. The practical part is prevention: wipe only with a well-wrung cloth, keep liquid away from vents and sensor channels, and follow the manufacturer's cleaning instructions.

Safety

Disconnect the machine from mains power before opening the pneumatic block; after any moisture-related repair, let it dry out completely and complete the electrical safety test and a full pneumatic leak test before returning it to clinical use.

Lesson: When every channel of one block misbehaves at the same time, look for a common cause reaching them all — such as moisture from cleaning — before replacing parts one by one.

Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes