Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Dräger Apollo (Авентa)
Dräger · Apollo (Авентa)

Dräger Apollo: set oxygen concentration not reached at low rail pressure

What happened

Reported in the case: with oxygen pressure in the rail reduced to about 1.2 atm, the machine generated no alarm but physically could not deliver the set oxygen concentration — the actual FiO2 was below the setting, and the oxygen sensor calibration would not pass.

What was checked

Cause

Probable cause per the case: low oxygen pressure in the rail combined with no low-pressure monitoring or alarm for this condition — reported for machines from the affected production period.

The cause is probable and was not fully confirmed.

What was done

No parts were replaced. The practical fix was to restore normal oxygen supply pressure to the machine; for on-site verification the engineers used their own oxygen cylinder as a known-good source. The engineers noted this on machines of the same type showed a similar picture, and the link between the low rail pressure and the symptoms remained probable rather than confirmed.

Safety

Before returning an anesthesia machine to use, verify the supply gas pressures and confirm the oxygen analyzer reads and alarms correctly against a gas source of known concentration — a reduced FiO2 without an alarm is a direct hypoxia risk to the patient.

Lesson: A missing alarm is not proof of a healthy gas supply — always verify supply pressure and delivered oxygen concentration from a known-good source before trusting the machine.

Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes