Dräger Avance: inspiration flow readings differ from identical units

What happened

Inspiration-side flow readings on one anaesthesia machine did not match identical units, raising doubts about the accuracy of the flow measurement on that machine.

What was checked

Cause

As reported in the case, the log timestamps showed the system tests (10:20) had been completed before the turbine calibration (10:47). Since the machine adapts its measurements to the circuit during the tests, the test results did not reflect the fresh calibration — a wrong order of calibration and tests rather than a faulty turbine, and most likely the cause of the discrepancy.

The cause is probable and was not fully confirmed.

What was done

In this case it was recommended to repeat the turbine calibration and then, in that order, run the full system tests with the working circuit fitted, confirming inspiration flow matches the comparable units; the outcome was reported as probable. Any calibration step behind service-level access is a procedure from the manufacturer's service documentation, done by authorised personnel.

Safety

Before the machine goes back into clinical use, run the complete system self-test and leak test with the circuit, absorber and vaporisers actually fitted, and check volume delivery on a test lung.

Lesson: Calibration and self-tests are order-dependent — after any calibration, re-run the full system tests, and check log timestamps before suspecting hardware when identical machines disagree.

Covered in the course: Ventilator Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.