Dräger Avena M (Carina): inspired–expired tidal volume gap
What happened
Measured inspiratory and expiratory tidal volumes diverged by more than 15%: at a set volume of 600 ml the display showed about 600 ml on inspiration but about 500 ml on expiration (figures as reported in the case). Standard measures did not shift the numbers.
What was checked
- Turbine and expiratory valve calibration was run per the manufacturer's service documentation — no effect on the gap.
- The flow sensor was swapped for a known-good one — numbers unchanged.
- The unit was tried on several different circuits and a test lung; the bag was replaced and the circuit checked for tightness — gap still there.
- Verification of the true delivered volume with an external flow analyser (a Citrex-type device) was proposed as the reference measurement.
Cause
No confirmed root cause in this case. The picture best fits a real leak in the breathing circuit that the ventilator was not compensating: inspired volume matched the setting while expired volume came back lower — so a sensor fault was unlikely.
The cause is probable and was not fully confirmed.
What was done
Engineers advised enabling leak compensation; the engineer in the case reacted positively after switching it on, but the final result was not confirmed in the discussion. Also recommended: confirm the true delivered volume with an external analyser before judging the unit. Bending the flow straightener in the inspiratory port was mentioned as an extra tweak — a non-standard measure, not a normal repair step.
Safety
Do not return the ventilator to clinical use while the volume gap is unexplained: verify accuracy on a test lung against an independent analyser and confirm the leak and disconnect alarms, because an uncompensated leak can silently under-ventilate a patient.
Lesson: A large inspired–expired volume gap is a leak symptom first — chase the leak and confirm delivered volume with an independent analyser before condemning sensors or valves.
Covered in the course: Ventilator Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.