Dräger Aventa-M: ventilation fault after user-replaced expiratory valve connector
What happened
Reported in the case: after a self-replacement of the expiratory valve connector and flow sensor, the machine stopped working correctly. The engineers discussed whether this replacement counts as a user-level procedure.
What was checked
- Remove the expiratory valve assembly and check the membrane orientation against the diagram in the documentation — in this case it had been refitted the wrong way round.
- Confirm the connector and flow sensor are fully seated, then run the machine check with a leak test and ventilation verification after any reassembly.
Cause
The engineers found the expiratory valve membrane had been reinstalled upside down during the earlier self-replacement; the machine itself was otherwise in order.
What was done
Turning the membrane to its correct orientation restored normal operation. The engineers noted the replacement is labour-intensive — disassembly covers roughly half the machine — but it is a user-accessible procedure described in the documentation.
Safety
Before an anaesthesia machine returns to patient use after any breathing-system work, run the full pre-use machine check and verify delivered volumes and pressures — a misassembled expiratory valve directly affects the ventilation the patient receives.
Lesson: When a machine misbehaves right after a part replacement, check assembly orientation against the documentation before suspecting a device fault — a flipped diaphragm can mimic one.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.