Dräger Caleo: humidity fault, evaporator block stuck on hidden latch
What happened
Humidity problem in the incubator; the engineer needed to remove the evaporator block but could not work out how to disassemble the assembly.
What was checked
- Reported in the case: two hex screws at the underside of the base were removed, but the block still would not come free, so the engineers looked for additional fasteners or latches.
- Inspection between the two screw heads showed a slot in the housing, which pointed to a hidden latch.
Cause
No root cause for the humidity fault was established in the case. The immediate obstacle was mechanical: apart from the two visible hex screws, the evaporator block is retained by an internal latch under the slot and by hooks on the opposite side, so it cannot be removed by unscrewing alone.
The cause is probable and was not fully confirmed.
What was done
As described in the case, access was purely mechanical: with the two hex screws out, a long screwdriver is inserted into the slot between them to press the latch aside; the lower floor panel is then lowered carefully, because its wiring harness has limited length; the connectors are disconnected, and the panel is lifted off the hooks on the opposite side without any screws. The reported outcome was probable — the disassembly procedure was shared in the case, with no confirmed repair result reported.
Safety
Isolate the incubator from mains power before opening the base, drain and disinfect the humidity system, and complete the electrical safety test before returning the unit to use, since standing water in the evaporator section is a biohazard risk for neonatal patients.
Lesson: When an assembly will not come free after all visible screws are out, expect hidden latches and hooks — look for release slots in the housing and check the harness length before pulling anything.
Covered in the course: Fault-Finding Method for Medical Equipment →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.