Dräger Evita XL: wheezing noise in the test lung during ventilation
What happened
During ventilation the test lung produced an audible wheeze. Its diaphragm and PEEP valve were intact, and the noise repeated with a different lung and with a bag, so the lung itself was not the culprit.
What was checked
- Swapped the test lung and tried a bag — the noise stayed.
- Inspected the diaphragm and PEEP valve of the test lung — both intact.
- Reviewed the set ventilation parameters, how AutoFlow behaved and whether the unit was riding up against the Ppeak limit.
- Varied PEEP and inspiratory time (Tinsp) to see what changed the noise.
Cause
Per the case, most likely a settings issue rather than a hardware defect: with AutoFlow active, PEEP together with a long inspiratory time pushed the breaths against the Ppeak limit, and the test lung answered with an audible wheeze. This is a probable explanation from one case, not a proven root cause.
The cause is probable and was not fully confirmed.
What was done
The engineers found that the wheeze disappeared when PEEP was set to 0 and Tinsp was reduced, as reported in the case; colleagues also suggested switching AutoFlow off. No parts were replaced — the ventilation settings were the fix (probable outcome).
Safety
Run all these checks on a test lung with the ventilator isolated from any patient circuit, and repeat the manufacturer's pre-use self-test before the unit returns to clinical use.
Lesson: A wheeze in the test lung is not always a broken lung or valve — walk the parameter set step by step first, because the settings themselves can be the fault.