Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Dräger Evita XL
Dräger · Evita XL

Dräger Evita XL: wheezing noise in the test lung during ventilation

What happened

During ventilation the test lung produced an audible wheeze. Its diaphragm and PEEP valve were intact, and the noise repeated with a different lung and with a bag, so the lung itself was not the culprit.

What was checked

Cause

Per the case, most likely a settings issue rather than a hardware defect: with AutoFlow active, PEEP together with a long inspiratory time pushed the breaths against the Ppeak limit, and the test lung answered with an audible wheeze. This is a probable explanation from one case, not a proven root cause.

The cause is probable and was not fully confirmed.

What was done

The engineers found that the wheeze disappeared when PEEP was set to 0 and Tinsp was reduced, as reported in the case; colleagues also suggested switching AutoFlow off. No parts were replaced — the ventilation settings were the fix (probable outcome).

Safety

Run all these checks on a test lung with the ventilator isolated from any patient circuit, and repeat the manufacturer's pre-use self-test before the unit returns to clinical use.

Lesson: A wheeze in the test lung is not always a broken lung or valve — walk the parameter set step by step first, because the settings themselves can be the fault.

Covered in the course: Ventilator Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes