Dräger Fabius GS: fan fault at ventilation start despite passed tests
What happened
As reported in the case, the ventilator passed all built-in tests and calibrations, but every time ventilation was started it threw a fan fault. From the service mode the fan could be driven manually and it passed the fan test. The engineers also noticed the interface language fell back to English after power-up — a sign that settings were not being retained.
What was checked
- Repeat the calibrations and the fan test, then power the unit off and on again and check whether the settings, options and calibrations survive the cycle.
- After each power-up, look at the interface language and configuration: booting with factory defaults points to a lost memory supply rather than a mechanical fault.
- Check the backup battery on the power/settings board: measure its voltage and confirm whether options and calibrations drop out at every power-on.
- If the settings hold, move to the fan itself: connectors, wiring and motor behaviour at start, since the fan passed tests while driven from service.
Cause
Suspected in this case, not confirmed: an exhausted backup battery on the power/settings board, so options and calibrations were lost at every power-up and the startup check flagged a fan fault. A worn fan motor was kept as the alternative until the battery is ruled out.
The cause is probable and was not fully confirmed.
What was done
Check the backup battery first; if it is flat, replace it and restore the machine configuration and calibrations — this is a procedure from the manufacturer's service documentation, done by authorised personnel. Then power-cycle the unit several times to confirm the settings hold and the fan fault does not return; if it does, test the fan and its wiring. The outcome in this case was probable, so confirm the fix before trusting the machine.
Safety
A ventilator that boots with default settings may not behave as expected; after any battery or configuration work, run the full electrical safety test and a complete pre-use check on a test lung before the unit returns to clinical use.
Lesson: When a device passes its tests but fails a check right after power-up, look at what resets between power cycles — a fallback language and default settings point to a dead memory battery, not a broken component.
Covered in the course: Ventilator Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.