Dräger Fabius: lost settings and multiple faults after power-up
What happened
After an overnight shutdown the machine powered up without its Russian-language interface, and faults appeared for the O2 sensor, the ventilator and the APL valve.
What was checked
- A batch of unrelated faults appearing together right after power-up, plus a lost interface language, pointed to loss of the settings memory rather than to individual component failures.
- The engineers checked the backup battery that powers the settings memory: it was completely flat, and the reset of date and time confirmed the memory had been unpowered.
Cause
The backup battery for the settings memory had discharged fully, so on the next start the machine came up with default date and time and had lost its calibrations, configuration and options — which surfaced as faults on the O2 sensor, the ventilator and the APL valve.
What was done
The engineers replaced the memory backup battery; after that the machine could be started in a single basic volume-controlled mode. Restoring the calibrations, settings and options was done as a procedure from the manufacturer's service documentation, performed by authorised personnel. The fault was cleared, as reported in the case.
Safety
A machine that has lost its settings must be treated as uncalibrated: do not return it to clinical use until all calibrations are redone and the full pre-use check, including O2 monitoring, ventilation and leak testing, passes.
Lesson: When a machine throws a group of unrelated faults immediately after being powered down, check the memory backup battery first before chasing individual components.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.