Dräger · Fabius plus

Dräger Fabius Plus: fails leak (tightness) test

What happened

The anesthesia machine does not pass its routine leak test; the leak could not be localised at first.

What was checked

Cause

The suspected cause, based on comparable cases, was a cracked tube on the line from the rotameter block; in this particular case the exact cause remained unconfirmed (probable).

The cause is probable and was not fully confirmed.

What was done

No completed fix was reported for this case. If the cracked tube is confirmed, the expected remedy is to replace that line with the correct spare part and then repeat the full leak test and the machine self-test; internal pneumatic work should follow the manufacturer's service documentation.

Safety

Depressurise and isolate the gas supplies before opening the pneumatic circuit, and before returning the machine to use run the full self-test plus a manual leak check.

Lesson: A whole-machine leak test failure is best narrowed down section by section — check the obvious seals first, but small hidden tubing such as the rotameter block line can be the culprit.

Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.