Dräger Fabius Plus: fails leak (tightness) test
What happened
The anesthesia machine does not pass its routine leak test; the leak could not be localised at first.
What was checked
- Inspect the breathing block valves and diaphragms for correct seating, cracks and wear.
- Check the absorber canister gasket — a tired or misplaced seal is a classic leak source.
- Confirm each vaporizer is fully seated and latched on its mount.
- Trace and inspect the small tube from the rotameter (flowmeter) block along its full length; look for hairline cracks, especially near fittings.
Cause
The suspected cause, based on comparable cases, was a cracked tube on the line from the rotameter block; in this particular case the exact cause remained unconfirmed (probable).
The cause is probable and was not fully confirmed.
What was done
No completed fix was reported for this case. If the cracked tube is confirmed, the expected remedy is to replace that line with the correct spare part and then repeat the full leak test and the machine self-test; internal pneumatic work should follow the manufacturer's service documentation.
Safety
Depressurise and isolate the gas supplies before opening the pneumatic circuit, and before returning the machine to use run the full self-test plus a manual leak check.
Lesson: A whole-machine leak test failure is best narrowed down section by section — check the obvious seals first, but small hidden tubing such as the rotameter block line can be the culprit.