Dräger Fabius Plus: flowmeter needle valve leak
What happened
As reported, an external leak at a flowmeter needle valve on a Dräger Fabius Plus anesthesia machine: gas wept past the closed valve.
What was checked
- The engineers found the valve from the flowmeter panel, behind the small plastic plugs marked with the gas symbol.
- They attempted to reseat the needle: loosen the retaining nut, then adjust the needle with a screwdriver; replacement of the needle valve was the fallback if the adjustment did not hold.
Cause
The case did not establish a definitive cause; the picture is consistent with the needle losing seat contact (wear or loosening), which let gas pass with the valve closed.
What was done
As reported in the case: the centre plug/cap was removed, the retaining nut loosened and the needle re-tightened with a screwdriver; if the adjustment does not hold, the needle valve is replaced. The leak was resolved, but the case did not confirm which step settled it — adjustment, replacement, or both. Work of this kind is done per the manufacturer's service documentation.
Safety
Isolate the machine from all gas supplies and depressurise before opening the valve, and before return to use verify leak-tightness and flow accuracy across the range — needle adjustment can shift the delivered flow.
Lesson: A weeping flowmeter valve often just needs the needle reseated, but any work on a gas-path valve counts as done only after a leak test and a flow check on the machine.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.