Dräger · Fabius Plus

Dräger Fabius Plus: flowmeter needle valve leak

What happened

As reported, an external leak at a flowmeter needle valve on a Dräger Fabius Plus anesthesia machine: gas wept past the closed valve.

What was checked

Cause

The case did not establish a definitive cause; the picture is consistent with the needle losing seat contact (wear or loosening), which let gas pass with the valve closed.

What was done

As reported in the case: the centre plug/cap was removed, the retaining nut loosened and the needle re-tightened with a screwdriver; if the adjustment does not hold, the needle valve is replaced. The leak was resolved, but the case did not confirm which step settled it — adjustment, replacement, or both. Work of this kind is done per the manufacturer's service documentation.

Safety

Isolate the machine from all gas supplies and depressurise before opening the valve, and before return to use verify leak-tightness and flow accuracy across the range — needle adjustment can shift the delivered flow.

Lesson: A weeping flowmeter valve often just needs the needle reseated, but any work on a gas-path valve counts as done only after a leak test and a flow check on the machine.

Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.