Dräger Fabius: preventive maintenance reminder stays on

What happened

The screen shows a 'preventive maintenance required' message, and the engineers wanted to clear it after servicing the machine.

What was checked

Cause

The message is a software service-interval reminder that switches on when the scheduled maintenance interval expires and stays on until it is cleared; in this case the engineers found it reportedly could not be reset on the installed software version, so a newer version was considered.

The cause is probable and was not fully confirmed.

What was done

As reported in the case, the probable fix was to update the software to a newer version (reported in the case as 3.37b), with a firmware re-flash also mentioned as an option, and then clear the reminder; both the update and the reset are procedures from the manufacturer's service documentation, done by authorised personnel, and the reminder is cleared only after the actual preventive maintenance has been completed.

Safety

Keep the anaesthesia machine out of clinical use until it has passed the full power-up self-test and a basic leak/ventilation check after the update, and disconnect it from the mains before any internal access.

Lesson: A maintenance reminder should be cleared only with the manufacturer's documented procedure by authorised personnel, and only after the preventive maintenance itself has actually been done.

Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.