Dräger · Fabius Tiro

Dräger Fabius Tiro: bellows jerks at startup, ventilator fault

What happened

On power-up the ventilator bellows jerks briefly and then does not move; the screen shows a ventilator fault.

What was checked

Cause

The exact fault was not confirmed in this case. The engineers found two likely contributors: the drive motor possibly not registering its lower limit position, and a flat CR2032 backup battery that let stored settings and options lapse — together this explains the single jerk at startup and the ventilator fault.

The cause is probable and was not fully confirmed.

What was done

In this case the plan was to replace the CR2032 backup battery with the machine kept connected to mains power so the settings are not lost, then run the pressure and PEEP calibrations; restoring the options was a procedure from the manufacturer's service documentation, done by authorised personnel. After that, volume mode was expected to be available again — the outcome was reported in the case as probable, not fully confirmed.

Safety

Because the backup battery is changed with the unit connected to mains, keep hands and tools clear of energised parts and complete an electrical safety check before the machine returns to clinical use.

Lesson: A dying backup battery can silently wipe stored settings and options on a ventilator, so check it before chasing motors and sensors.

Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.