Dräger Fabius Tiro: compliance and tightness test fails with no visible leak
What happened
On a Dräger Fabius Tiro anesthesia machine, the automatic compliance and leak test kept failing even though no visible leak could be found anywhere in the breathing system.
What was checked
- Swap the breathing circuit for a known-good one and repeat the test to rule out the hose set itself.
- Check that the vaporiser filler caps are fully closed and that the vaporiser is not passing gas or agent.
- Inspect the O-rings on the O2 sensor connector and on any plugged ports for flattening, drying or cracks.
- Wipe the seals and mating surfaces with an alcohol wipe and retest; dirt on an O-ring can leak without being visible.
- If the test still fails, move on to the inspiratory and expiratory ceramic valves and the O2 sensor.
Cause
No definitive cause was established in this case. The engineers judged the most likely culprits to be small, invisible leaks: tired or dirty O-rings on the O2 sensor connector and plugs, a leaking vaporiser, or internal leakage past the inspiratory and expiratory ceramic valves.
The cause is probable and was not fully confirmed.
What was done
The fix reported in the case was to work through the leak points in order: swap in a known-good breathing circuit, clean the seals, replace suspect O-rings (especially at the O2 sensor connector and plugs), verify the vaporiser caps and the vaporiser itself, and, if the test still fails, inspect the inspiratory and expiratory ceramic valves. The outcome was probable: the checks narrowed the fault to this shortlist, but the final root cause was never confirmed.
Safety
Do not return the machine to clinical use until the compliance and tightness test passes completely — an unnoticed leak can distort delivered volumes, pressures and the gas and agent mixture the patient receives.
Lesson: When a tightness test fails with no visible leak, work from the simple and likely to the internal: seals, O-rings and plugs first, valves last — a small invisible leak is still a leak.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.