Dräger Fabius Tiro: flow sensor fails calibration, volume not displayed
What happened
On a Dräger Fabius Tiro anesthesia machine the flow sensor would not pass calibration and the measured volume was not displayed.
What was checked
- Visual inspection of the flow sensor for damage and contamination
- Continuity check of the sensor with a multimeter, which showed an open circuit in one conductor
Cause
An open circuit in one of the conductors inside the flow sensor, as reported in the case; the machine itself was not at fault.
What was done
The flow sensor was replaced. After replacement the calibration passed and the volume display returned. The engineers noted that these sensors are wear items, so replacement rather than repair is the expected remedy.
Safety
Before returning an anesthesia machine to clinical use after flow-sensor work, run the machine's complete self-test and verify measured volumes against a reference per the manufacturer's service documentation.
Lesson: When a flow sensor fails calibration, test the sensor itself first — a simple continuity check can reveal an internal wiring fault in a worn expendable part.