Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Dräger Fabius Tiro
Dräger · Fabius Tiro

Dräger Fabius Tiro: flow sensor fails calibration, volume not displayed

What happened

On a Dräger Fabius Tiro anesthesia machine the flow sensor would not pass calibration and the measured volume was not displayed.

What was checked

Cause

An open circuit in one of the conductors inside the flow sensor, as reported in the case; the machine itself was not at fault.

What was done

The flow sensor was replaced. After replacement the calibration passed and the volume display returned. The engineers noted that these sensors are wear items, so replacement rather than repair is the expected remedy.

Safety

Before returning an anesthesia machine to clinical use after flow-sensor work, run the machine's complete self-test and verify measured volumes against a reference per the manufacturer's service documentation.

Lesson: When a flow sensor fails calibration, test the sensor itself first — a simple continuity check can reveal an internal wiring fault in a worn expendable part.

Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes