Dräger Fabius Tiro: O2 sensor replacement and calibration

What happened

Question on the correct procedure for replacing the oxygen sensor and calibrating the new one.

What was checked

Cause

No fault as such: the galvanic O2 cell is a consumable with a limited service life and was due for replacement.

What was done

Fit the stabilised sensor, calibrate it against room air and confirm there are no error messages. In this case the engineer first reset the amplifier channel offsets via the service menu — that was a procedure from the manufacturer's service documentation, done by authorised personnel. A second opinion in the case held that room-air calibration alone was enough and the offset reset may not always be needed; that was not confirmed in one case. As reported, removing the sensor from the breathing circuit to ambient air during downtime extends its life.

Safety

Before returning the machine to use, verify the O2 reading against a gas of known concentration and confirm the low-oxygen alarm triggers — unverified oxygen monitoring during anaesthesia is a hypoxia risk.

Lesson: Treat the oxygen sensor as a consumable: let a new cell stabilise in air, calibrate against room air, leave service-level steps to authorised personnel, and store the cell outside the circuit when the machine is idle.