Dräger · Fabius

Dräger Fabius: worn dual pneumatic hose (cuff line)

What happened

On a Dräger Fabius anesthesia machine the dual pneumatic hose serving the cuff line needed replacement; the engineers were looking for the original part or a suitable substitute.

What was checked

Cause

The case did not state the exact failure mode; per the engineers the hose was simply due for replacement. Typical causes with this part are age hardening, cracking and kinking at the bends.

What was done

In this case the engineers replaced the dual hose with white silicone hose reported as a carburetor-type analog from an automotive shop; a suitable hose also exists in the manufacturer's service kit, and coils reported in the case as about 30 m can be bought on an online marketplace and cut to the required length. Outcome: solved. The substitute must be clean, flexible, material-compatible silicone with matching dimensions and secure clamping; the service-kit part remains the first choice when available.

Safety

Since this line is part of the patient pneumatic circuit, before return to use perform a full leak test, the machine's system self-test and a functional check of cuff inflation and pressure; use only clean hose made of material suitable for medical use.

Lesson: A generic silicone hose can bridge the gap when the original dual hose is unavailable, but only after verifying dimensions, material suitability and a proven leak-free result.

Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.