Dräger GS: no 24 V, mechanics dead
What happened
The machine lost its 24 V rail, the mechanical ventilator stopped moving, and battery charging was out as well.
What was checked
- Check for 24 V at the loads first, to confirm the whole rail is down rather than a single branch.
- Trace the rail back to its source and measure the power supply output; as reported in the case, it should read about 28 V on this unit.
- Remove the rear panel and inspect the power supply board: damaged components, and trimmer potentiometers that look seized, corroded or heavily marked by a screwdriver.
Cause
As reported, the trimmer potentiometers on the power supply had seized and lost wiper contact, which took down the 24 V rail together with the battery charging circuit.
The cause is probable and was not fully confirmed.
What was done
The engineers gently worked the seized trimmers back and forth to restore contact; as reported, the 24 V rail returned and battery charging resumed. Since output trimmers also set the voltage, the PSU outputs (about 28 V as reported) were to be re-checked against the service documentation, and any voltage re-adjustment is a procedure from the manufacturer's service documentation, done by authorised personnel. The outcome was probable, not fully verified.
Safety
Unplug the machine and let the power supply capacitors discharge before opening the PSU, and complete the electrical safety test and full function check — including mains charging of the battery — before the anesthesia machine is returned to patient use.
Lesson: A completely dead 24 V rail can come down to a single dried-out trimmer contact in the power supply, so measure the source before swapping loads or modules.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.