Dräger GS: no 24 V, mechanics dead

What happened

The machine lost its 24 V rail, the mechanical ventilator stopped moving, and battery charging was out as well.

What was checked

Cause

As reported, the trimmer potentiometers on the power supply had seized and lost wiper contact, which took down the 24 V rail together with the battery charging circuit.

The cause is probable and was not fully confirmed.

What was done

The engineers gently worked the seized trimmers back and forth to restore contact; as reported, the 24 V rail returned and battery charging resumed. Since output trimmers also set the voltage, the PSU outputs (about 28 V as reported) were to be re-checked against the service documentation, and any voltage re-adjustment is a procedure from the manufacturer's service documentation, done by authorised personnel. The outcome was probable, not fully verified.

Safety

Unplug the machine and let the power supply capacitors discharge before opening the PSU, and complete the electrical safety test and full function check — including mains charging of the battery — before the anesthesia machine is returned to patient use.

Lesson: A completely dead 24 V rail can come down to a single dried-out trimmer contact in the power supply, so measure the source before swapping loads or modules.

Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.