Academy / Field fault notes / Dräger Leon
Dräger · Leon

Dräger Leon: zero PEEP and abnormal ventilation

What happened

The ventilator behaved incorrectly during use, and PEEP stayed at zero as reported in the case, so a fault in the machine itself was suspected.

What was checked

Cause

The engineers found that the rubber membrane of the expiratory valve had been fitted flipped, in the wrong orientation.

What was done

Re-seat the expiratory valve membrane in its correct orientation, then confirm normal operation on a test lung and complete the manufacturer's pre-use verification before returning the ventilator to service.

Safety

Never troubleshoot a ventilator on a patient: disconnect it and work on a test lung, and make sure it passes the manufacturer's pre-use verification before it goes back into service.

Lesson: A zero reading on a basic parameter like PEEP can come from a simply misassembled valve membrane, so check the seating and orientation of freshly serviced parts before suspecting electronics.

Covered in the course: Ventilator Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.