Dräger Primus: "Check fan assembly" message although circuit and self-test are OK
What happened
A Dräger Primus anesthesia machine kept showing a "Check fan assembly" message even though the breathing circuit had been re-assembled, the bellows looked intact and the machine self-test passed.
What was checked
- The engineers re-assembled the breathing circuit, inspected the bellows and ran the machine self-test; the message still appeared.
- They checked whether the gas analyzer sampling tube was properly seated on the circuit tee.
- They also checked the flow sensor connectors on the LP Analog board for being swapped.
Cause
The final cause was not confirmed in this case. Two probable causes were proposed: a disconnected gas analyzer sampling tube at the tee, and the flow sensor connectors swapped on the LP Analog board — either can distort the flow measurement enough to trigger the fan assembly message.
The cause is probable and was not fully confirmed.
What was done
Probable fix, not confirmed in this case: reconnect the gas analyzer sampling tube at the tee, re-seat the flow sensor connectors on the LP Analog board in their correct positions, and repeat the self-test.
Safety
Unplug the machine from mains power before removing covers to reach internal boards, and after any work on the breathing system run the full self-test plus a test-lung ventilation check before the machine returns to clinical use.
Lesson: Before blaming the fan or bellows, walk the whole flow-measurement chain — sampling lines and sensor connectors can mimic a mechanical assembly fault.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.