Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Dräger Primus
Dräger · Primus

Dräger Primus: elevated inspired CO2 on the breathing circuit

What happened

Elevated inspired CO2 (7–10 mmHg as reported in the case), with initial suspicion of a leak from the expiratory to the inspiratory side past the CO2 absorber.

What was checked

Cause

The engineers found that the lime from one specific canister was of poor quality or exhausted despite looking fresh, so CO2 passed through without being fully absorbed; the suspected internal leak past the absorber was not confirmed.

What was done

The soda lime was replaced with lime from a different canister, and the inspired CO2 readings returned to normal; the questionable canister was taken out of use.

Safety

When changing absorber lime, wear eye protection and a dust mask, make sure the canister is seated correctly, and run the full pre-use leak and performance check, including inspired-CO2 verification, before returning the machine to service.

Lesson: Even fresh-looking soda lime can be from a bad batch, so when inspired CO2 is high, prove the absorbent itself (for example by swapping canisters) before chasing leaks or blaming the analyzer.

Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes