Dräger Primus: switches to battery during boot-up after a mains surge
What happened
After a voltage spike in the mains, the machine powers on showing mains present, but during boot-up, before the self-test, it switches itself to battery even though the wall supply is still connected.
What was checked
- The engineers inspected the power supply start-related connections and mains-side wiring; nothing conclusive was reported.
- As done in the case, the machine was left connected to mains with the battery fitted for a full day so the system could re-evaluate the battery.
Cause
Not finally confirmed. The working assumption in the case: the surge upset the battery evaluation, and the Primus power supply does not start properly from mains with a deeply discharged or unrecognised battery, so the unit falls back to battery during boot.
The cause is probable and was not fully confirmed.
What was done
In this case the engineers left the machine plugged into mains with the battery connected for 24 hours (as reported in the case) so the system could re-analyse the battery, then repeated the start-up check. The outcome was described as probable, not confirmed — afterwards, verify the actual battery condition and the mains start behaviour before trusting the machine.
Safety
Before returning the machine to clinical use, run the electrical safety test and prove the battery backup holds the machine through a mains-failure check — an anaesthesia machine that drops onto a weak battery when mains disappears is a direct patient hazard.
Lesson: After a mains surge on an anaesthesia machine, suspect the battery state first and give the system time to re-evaluate it before tearing into the power supply.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.