Dräger Primus: will not power up with depleted batteries
What happened
A Dräger Primus anaesthesia machine that would not power on; the discussion also covered whether the unit can run at all with faulty rechargeable batteries and which software version is current.
What was checked
- Check the external power path first — mains cord, inlet and the power supply output — as the engineers did in this case, so a supply fault is not mistaken for dead batteries.
- Test each battery pack on its own: voltage at rest and under load; packs reported in the case as dead kept the machine from starting even with mains connected.
- After battery replacement, leave the machine powered and observe whether the audible alarm clears by itself within the time reported in the case, instead of immediately chasing a fault.
Cause
Reported cause: fully discharged or faulty rechargeable batteries block power-up — the engineers found that in this case the batteries were not merely a lost backup but the reason the unit would not start; a power supply fault was considered and ruled out.
The cause is probable and was not fully confirmed.
What was done
The fix reported in the case was replacement of the battery packs: the engineers' position was that the machine needs serviceable batteries to power on, so dead packs must be replaced, not bypassed. They also reported that after the swap the audible warning can persist for roughly 15 minutes (as stated in the case) and then clears on its own provided the power supply is healthy. The software version named in the case as current was 4.53.01. Note that the outcome was reported as probable, not fully confirmed.
Safety
Before returning the machine to use, isolate it from mains before handling the batteries, and afterwards verify correct operation on both mains and battery power together with the electrical safety checks required before clinical use.
Lesson: On an anaesthesia machine the batteries are part of the power-on path, not just an emergency backup, so test them first whenever a unit refuses to start.
Covered in the course: Anaesthesia Machine Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.