Dräger Savina 300: O2 sensor voltage out of range
What happened
On a Dräger Savina 300, one of the oxygen sensors showed an output voltage outside the working range reported in the case (0.5–1.0 V). The engineers discussed how long these sensors actually last and when to replace them.
What was checked
- Measure the output voltage of each oxygen sensor separately and compare it with the working range reported in the case (0.5–1.0 V).
- Check the machine's operating hours and the oxygen concentrations it mostly ran at — electrochemical O2 cells age faster with high oxygen exposure.
Cause
The engineers concluded that one oxygen sensor had most likely simply degraded with age: its output drifted beyond the normal range.
The cause is probable and was not fully confirmed.
What was done
The engineers recommended replacing the oxygen sensors as a pair so both cells stay matched. Replacement intervals reported in the case varied from 3 to 6 years, though the same sensors often last longer when the ventilator runs mostly at low oxygen concentrations. Fit approved sensors and calibrate them according to the manufacturer's manual.
Safety
After replacing an oxygen sensor, run the sensor calibration and verify the displayed oxygen concentration against an independent analyzer before returning the ventilator to use — an aged or incorrectly set sensor can silently show the wrong FiO2.
Lesson: Electrochemical oxygen sensors are consumables that age with operating hours and oxygen exposure — track their output voltage over time and replace them before they drift out of range, not after.
Covered in the course: Ventilator Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.